Share Categories Channel
1/6
ITG O2 Oxygen Sensor Medical Sensor 0-100 Vol% O2/M-03, O2 Sensor, Oxygen Sensor image 1ITG O2 Oxygen Sensor Medical Sensor 0-100 Vol% O2/M-03, O2 Sensor, Oxygen Sensor image 2ITG O2 Oxygen Sensor Medical Sensor 0-100 Vol% O2/M-03, O2 Sensor, Oxygen Sensor image 3ITG O2 Oxygen Sensor Medical Sensor 0-100 Vol% O2/M-03, O2 Sensor, Oxygen Sensor image 4ITG O2 Oxygen Sensor Medical Sensor 0-100 Vol% O2/M-03, O2 Sensor, Oxygen Sensor image 5ITG O2 Oxygen Sensor Medical Sensor 0-100 Vol% O2/M-03, O2 Sensor, Oxygen Sensor image 6

ITG O2 Oxygen Sensor Medical Sensor 0-100 Vol% O2/M-03, O2 Sensor, Oxygen Sensor

Advanced Sensors Application Technology Co., Ltd.

Price $12.00 / Piece
Mini Order
-+Subtotal: $12
Availability In Stock
Purchase details

Payments:

Returns&Refunds:

Eligible for returns and refunds View More

Model NO.
O2/M-03
Type
Electrochemical Gas Sensor
Conductivity Type
Volume Control Type
Output Signal Type
Analog Type
Production Process
Integration
Material
PA, PPS, PTFE, Stainless Steel
IP Rating
IP20
Sensitivity
High
Certification
RoHS
Customized
Customized
Output
TTL/4-20mA/RS485
Transmitter Board
Integrated/Seperated
Wire
Two or Three Wire
Detecting Range
Selectable
After Sale Service
Yes
Trademark
ITG
Transport Package
Factory Sealed
Specification
29.2*41.4
Origin
Germany
HS Code
9027100090
Sensor Key Feature

Long life gas sensor that shows high signal stability at low cross interferences to anesthesia gases combined with superior linearity over the entire range.


All characteristics are based on conditions at 25ºC, 50% RH and 1013 hPa.
Measurement Range0 to 100 Vol.%
Expected Operating Life~ 1,000,000 Vol.% h
Sensor Lifetime< 6 years @ ambient air, depending on application
Electrical Connector3mm mono jack
Initial Output Signal13.0 to 16.5 mV @ dry ambient air
Response Time t90< 12 s
Drift< 1 % Vol. O2/month@ air, averaged across 12 months
Operating Temperature10 to 40 °C
Pressure Range700 to 1250 hPa
Linearity Error≤3@100 % O2, applied for 5 min
Zero Offset Voltage≤0.3 Vol O2@100% N2 applied for 5 min
Repeatability± 1 % Vol. O2 @ 100 Vol % O2 applied for 5 min
Influence of Humidity- 0.03 % rel. O2 reading per % RH
Recommended Load Resistor≥ 10 kOhm
Temperature CompensationNTC
Interferencesaccording to DIN EN ISO 80601-2-55
Weightapproximately 26 g
Material in Contact with MediaPA, PPS, PTFE, stainless steel

Sensor Storage Conditions
Temperature Rangerecommended:  5 to 30 °C
maximum:  -20 to 50 °C (≤10h)
Humidityup to 100 % RH, non-condensing
Shelf Life< 6 months recommended


Oxygen Module Parameters
Module typeGM-I2(3)-M/C/S-**-****/G/P/F/FL
Detected gasToxic gas
Detecting principleElectrochemistry
SamplingDiffusing  pump  flange  thread
Responding time30s~40s(of the sensor)
Module accuracy≤±3%FS(actual concentration, higher accuracy depends on specified sensors)
Module repeatability≤±2% FS
Module linear error≤±2% FS
Module zero temperature drift≤±2% (FS/year)
Module operating temperature-40~80°C
Module humidity0-95%RH(non-condensation)
Sensor operating temperature-20°C~50°C
Sensor humidity0-90%RH(non-condensation)
Signal outputTTL/4-20mA /RS485   (two or three wire)
Transmitting distance1000 meters(RS485)
Operating voltageDC 12-36V
Maximum power≤0.8W
Size80*48mm
Explosion proofIntrinsic safe
Connection cableRVVP3 x 1.5mm²

Oxygen Module Applications
Petroleum, petrochemical, metallurgy, chemical, electricity, telecom, gas, mining, pharmacy, municipal wastewater treatment, indoor environment, atmospheric environmental protection, transit equipment, transportation facilities, tunnel safety and underground parking, leak detection, safety and environmental control


 

Send Message